4. Instrument Setup
4.8 Filling and inserting the magazines
Fig. 25
• Filloneofthemagazines(48) with slides (49).
While doing so, ensure that the surface to be
imprinted (50) is on the left and facing upwards
(Fig. 23).
48
49
50
51
52
• Sothatthespecimenslidesareejectedcor-
rectly, ensure that the specimen slides are
inserted into the magazine accurately aligned
above each other (Fig. 26) and at a right angle
to the aperture (Fig. 27).
• Holdmagazine(48) at a slight angle (Fig. 24)
to prevent the specimen slides from falling
out of the magazine. Insert magazine into the
corresponding receptacle as shown. Both pins
(52) of the magazine need to lock into slot (51).
Fig. 23
Fig. 24
Fig. 26
Fig. 27